ProCide PA provides the user with fast, efficient and safe handling and is one of the best choices for highlevel disinfection of medical instruments and endoscopes. The aldehyde-free formulation makes the product safer for the environment and users. ProCide PA has a wide material compatibility when used at recommended concentrations and durations of use.
INSTRUCTIONS FOR USE
Broad spectrum of efficacy from 30 minutes
SIMPLEBefore using ProCide PA, clean the instrument surfaces and canals with a suitable enzymatic cleaner. (We recommend ProZyme +5). After cleaning, rinse and dry the instruments. Manual usage: First fill the disinfection bath with water (at room temperature), add the product at an appropriate amount, wait for 10-15 seconds. Make sure that all instruments and all canals come into contact with the solution during application. The lid of the disinfectant bath must be kept closed during use. The noncontaminated solution is stable for 7 days. The minimum effective concentration (MEC) should be checked with the test strips. If the color in test strips changes, the solution should be changed. After the recommended contact time, the medical devices should be rinsed with water of at least drinking water quality and dried.
Usage with the endoscope washing machine: Always follow the device manufacturer’s instructions for use. The product can be used with automatic dosing by connecting to the disinfectant pump of endoscope washing machines. Recommended concentration of use for high-level disinfection: 3% – 10 minutes.